| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.776 | 2.712 |
| 10.00 | 1.652 | 1.588 |
| 5.00 | 0.962 | 0.898 |
| 2.50 | 0.546 | 0.482 |
| 1.25 | 0.269 | 0.205 |
| 0.63 | 0.120 | 0.056 |
| 0.32 | 0.082 | 0.018 |
| 0.00 | 0.064 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-88% | 84% |
| EDTA plasma(n=5) | 92-99% | 96% |
| Heparin plasma(n=5) | 89-101% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-105% | 90-98% | 87-94% | 86-99% |
| EDTA plasma(n=5) | 84-90% | 79-93% | 83-93% | 83-93% |
| Heparin plasma(n=5) | 94-103% | 87-100% | 84-94% | 79-89% |