| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 10000.00 | 0.110 | |
| 5000.00 | 0.231 | |
| 2500.00 | 0.391 | |
| 1250.00 | 0.700 | |
| 625.00 | 1.065 | |
| 312.50 | 1.704 | |
| 156.25 | 1.822 | |
| 0.00 | 2.263 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-97% | 94% |
| EDTA plasma(n=5) | 94-108% | 101% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 77-83% | 81-88% | 95-101% | 96-104% |
| EDTA plasma(n=5) | 81-92% | 77-90% | 96-105% | 96-105% |
| Heparin plasma(n=5) | 84-90% | 89-97% | 85-95% | 82-93% |