| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 3.355 | 3.295 |
| 2500.00 | 1.847 | 1.787 |
| 1250.00 | 1.167 | 1.107 |
| 625.00 | 0.552 | 0.492 |
| 312.50 | 0.321 | 0.261 |
| 156.25 | 0.162 | 0.102 |
| 78.13 | 0.099 | 0.039 |
| 0.00 | 0.060 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-100% | 94% |
| EDTA plasma(n=5) | 81-90% | 86% |
| Heparin plasma(n=5) | 77-89% | 83% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-94% | 96-103% | 95-105% | 92-100% |
| EDTA plasma(n=5) | 96-107% | 79-92% | 86-98% | 86-98% |
| Heparin plasma(n=5) | 93-101% | 86-99% | 95-103% | 93-103% |