| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.614 | 2.555 |
| 1000.00 | 1.553 | 1.494 |
| 500.00 | 0.832 | 0.773 |
| 250.00 | 0.518 | 0.459 |
| 125.00 | 0.339 | 0.280 |
| 62.50 | 0.171 | 0.112 |
| 31.25 | 0.086 | 0.027 |
| 0.00 | 0.059 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-99% | 95% |
| EDTA plasma(n=5) | 87-101% | 94% |
| Heparin plasma(n=5) | 96-105% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-86% | 81-94% | 89-97% | 83-94% |
| EDTA plasma(n=5) | 87-98% | 95-102% | 87-94% | 87-94% |
| Heparin plasma(n=5) | 87-98% | 95-103% | 77-89% | 83-96% |