| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.331 | 2.254 |
| 2000.00 | 1.899 | 1.822 |
| 1000.00 | 1.044 | 0.967 |
| 500.00 | 0.573 | 0.496 |
| 250.00 | 0.326 | 0.249 |
| 125.00 | 0.180 | 0.103 |
| 62.50 | 0.120 | 0.043 |
| 0.00 | 0.077 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-87% | 82% |
| EDTA plasma(n=5) | 81-87% | 84% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 77-91% | 78-87% | 94-106% | 91-104% |
| EDTA plasma(n=5) | 81-95% | 89-103% | 89-97% | 89-97% |
| Heparin plasma(n=5) | 96-106% | 86-97% | 94-105% | 84-90% |