| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.725 | 2.660 |
| 1000.00 | 1.670 | 1.605 |
| 500.00 | 0.998 | 0.933 |
| 250.00 | 0.466 | 0.401 |
| 125.00 | 0.240 | 0.175 |
| 62.50 | 0.127 | 0.062 |
| 31.25 | 0.078 | 0.013 |
| 0.00 | 0.065 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-105% | 100% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 86-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-85% | 77-88% | 83-97% | 81-90% |
| EDTA plasma(n=5) | 92-104% | 86-99% | 86-100% | 86-100% |
| Heparin plasma(n=5) | 83-89% | 90-97% | 81-95% | 87-94% |