| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (mIU/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.227 | 2.174 |
| 25.00 | 1.452 | 1.399 |
| 12.50 | 0.927 | 0.874 |
| 6.25 | 0.443 | 0.390 |
| 3.13 | 0.274 | 0.221 |
| 1.57 | 0.156 | 0.103 |
| 0.79 | 0.118 | 0.065 |
| 0.00 | 0.053 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-99% | 94% |
| EDTA plasma(n=5) | 90-104% | 97% |
| Heparin plasma(n=5) | 92-102% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-104% | 79-89% | 91-104% | 82-95% |
| EDTA plasma(n=5) | 80-89% | 88-97% | 82-91% | 82-91% |
| Heparin plasma(n=5) | 79-85% | 78-92% | 91-104% | 88-98% |