| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 3.378 | 3.310 |
| 5.00 | 2.602 | 2.534 |
| 2.50 | 1.466 | 1.398 |
| 1.25 | 0.730 | 0.662 |
| 0.63 | 0.434 | 0.366 |
| 0.32 | 0.233 | 0.165 |
| 0.16 | 0.170 | 0.102 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-104% | 100% |
| EDTA plasma(n=5) | 96-106% | 101% |
| Heparin plasma(n=5) | 84-92% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-96% | 80-94% | 86-97% | 77-86% |
| EDTA plasma(n=5) | 83-89% | 91-98% | 96-107% | 96-107% |
| Heparin plasma(n=5) | 93-106% | 79-89% | 93-105% | 89-102% |