| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.373 | 2.322 |
| 50.00 | 1.720 | 1.669 |
| 25.00 | 1.118 | 1.067 |
| 12.50 | 0.752 | 0.701 |
| 6.25 | 0.377 | 0.326 |
| 3.13 | 0.229 | 0.178 |
| 1.57 | 0.116 | 0.065 |
| 0.00 | 0.051 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-102% | 96% |
| EDTA plasma(n=5) | 92-104% | 98% |
| Heparin plasma(n=5) | 80-94% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-104% | 79-91% | 78-92% | 86-99% |
| EDTA plasma(n=5) | 86-95% | 87-100% | 91-100% | 91-100% |
| Heparin plasma(n=5) | 96-105% | 84-90% | 92-99% | 78-86% |