| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 3.361 | 3.301 |
| 5.00 | 2.378 | 2.318 |
| 2.50 | 1.599 | 1.539 |
| 1.25 | 0.850 | 0.790 |
| 0.63 | 0.464 | 0.404 |
| 0.32 | 0.241 | 0.181 |
| 0.16 | 0.132 | 0.072 |
| 0.00 | 0.060 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-101% | 96% |
| EDTA plasma(n=5) | 93-106% | 100% |
| Heparin plasma(n=5) | 91-100% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 91-104% | 81-92% | 79-86% |
| EDTA plasma(n=5) | 88-95% | 88-101% | 80-89% | 80-89% |
| Heparin plasma(n=5) | 92-106% | 95-101% | 77-88% | 94-104% |