| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.812 | 2.727 |
| 25.00 | 1.926 | 1.841 |
| 12.50 | 1.126 | 1.041 |
| 6.25 | 0.576 | 0.491 |
| 3.13 | 0.295 | 0.210 |
| 1.57 | 0.179 | 0.094 |
| 0.79 | 0.122 | 0.037 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-88% | 84% |
| EDTA plasma(n=5) | 93-107% | 100% |
| Heparin plasma(n=5) | 84-90% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 92-102% | 83-89% | 91-98% |
| EDTA plasma(n=5) | 87-95% | 82-89% | 87-94% | 87-94% |
| Heparin plasma(n=5) | 83-94% | 93-102% | 77-84% | 79-90% |