| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 3.358 | 3.308 |
| 500.00 | 2.673 | 2.623 |
| 250.00 | 1.777 | 1.727 |
| 125.00 | 0.937 | 0.887 |
| 62.50 | 0.397 | 0.347 |
| 31.25 | 0.221 | 0.171 |
| 15.63 | 0.108 | 0.058 |
| 0.00 | 0.050 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-102% | 98% |
| EDTA plasma(n=5) | 80-88% | 84% |
| Heparin plasma(n=5) | 87-93% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-93% | 96-107% | 79-91% | 93-104% |
| EDTA plasma(n=5) | 80-91% | 84-92% | 77-86% | 77-86% |
| Heparin plasma(n=5) | 89-100% | 77-83% | 86-98% | 88-98% |