| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.473 | 2.398 |
| 25.00 | 1.864 | 1.789 |
| 12.50 | 1.121 | 1.046 |
| 6.25 | 0.685 | 0.610 |
| 3.13 | 0.361 | 0.286 |
| 1.57 | 0.199 | 0.124 |
| 0.79 | 0.122 | 0.047 |
| 0.00 | 0.075 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-91% | 84% |
| EDTA plasma(n=5) | 94-102% | 98% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-100% | 85-95% | 84-90% | 79-88% |
| EDTA plasma(n=5) | 79-89% | 79-85% | 86-96% | 86-96% |
| Heparin plasma(n=5) | 79-86% | 88-101% | 94-107% | 88-95% |