| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 3.069 | 3.004 |
| 5.00 | 2.059 | 1.994 |
| 2.50 | 1.210 | 1.145 |
| 1.25 | 0.757 | 0.692 |
| 0.63 | 0.429 | 0.364 |
| 0.32 | 0.267 | 0.202 |
| 0.16 | 0.164 | 0.099 |
| 0.00 | 0.065 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-83% | 80% |
| EDTA plasma(n=5) | 96-104% | 100% |
| Heparin plasma(n=5) | 87-94% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-104% | 82-90% | 92-98% | 93-102% |
| EDTA plasma(n=5) | 95-105% | 86-92% | 94-103% | 94-103% |
| Heparin plasma(n=5) | 82-90% | 92-101% | 77-89% | 91-100% |