| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.291 | 2.206 |
| 5.00 | 1.597 | 1.512 |
| 2.50 | 1.157 | 1.072 |
| 1.25 | 0.826 | 0.741 |
| 0.63 | 0.539 | 0.454 |
| 0.32 | 0.315 | 0.230 |
| 0.16 | 0.189 | 0.104 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-103% | 100% |
| EDTA plasma(n=5) | 92-100% | 96% |
| Heparin plasma(n=5) | 95-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-100% | 96-102% | 80-88% | 84-90% |
| EDTA plasma(n=5) | 77-89% | 78-89% | 93-106% | 93-106% |
| Heparin plasma(n=5) | 88-98% | 95-107% | 84-90% | 78-86% |