| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.027 | 1.954 | 
| 2500.00 | 1.497 | 1.424 | 
| 1250.00 | 1.069 | 0.996 | 
| 625.00 | 0.883 | 0.810 | 
| 312.50 | 0.559 | 0.486 | 
| 156.25 | 0.303 | 0.230 | 
| 78.13 | 0.232 | 0.159 | 
| 0.00 | 0.073 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-105% | 97% | 
| EDTA plasma(n=5) | 80-95% | 87% | 
| Heparin plasma(n=5) | 86-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-97% | 88-95% | 92-101% | 87-98% | 
| EDTA plasma(n=5) | 92-101% | 82-98% | 89-97% | 85-93% | 
| Heparin plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |