| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.354 | 2.260 |
| 2500.00 | 1.527 | 1.433 |
| 1250.00 | 1.269 | 1.175 |
| 625.00 | 0.876 | 0.782 |
| 312.50 | 0.593 | 0.499 |
| 156.25 | 0.327 | 0.233 |
| 78.13 | 0.236 | 0.142 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 82-94% | 85-96% | 78-90% |
| EDTA plasma(n=5) | 79-94% | 87-96% | 78-93% | 82-90% |
| Heparin plasma(n=5) | 86-102% | 87-99% | 99-105% | 95-104% |