| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 3.194 | 3.130 |
| 25.00 | 2.425 | 2.361 |
| 12.50 | 1.419 | 1.355 |
| 6.25 | 0.801 | 0.737 |
| 3.13 | 0.447 | 0.383 |
| 1.57 | 0.221 | 0.157 |
| 0.79 | 0.133 | 0.069 |
| 0.00 | 0.064 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-104% | 99% |
| EDTA plasma(n=5) | 94-100% | 97% |
| Heparin plasma(n=5) | 93-106% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-101% | 84-95% | 90-98% | 79-89% |
| EDTA plasma(n=5) | 94-102% | 78-84% | 88-102% | 88-102% |
| Heparin plasma(n=5) | 92-100% | 77-90% | 80-91% | 84-97% |