| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 20.00 | 2.195 | 2.108 | 
| 10.00 | 1.657 | 1.570 | 
| 5.00 | 1.263 | 1.176 | 
| 2.50 | 0.864 | 0.777 | 
| 1.25 | 0.509 | 0.422 | 
| 0.63 | 0.366 | 0.279 | 
| 0.32 | 0.258 | 0.171 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-92% | 85% | 
| EDTA plasma(n=5) | 80-97% | 88% | 
| Heparin plasma(n=5) | 88-102% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 91-99% | 88-97% | 88-104% | 97-108% | 
| EDTA plasma(n=5) | 83-97% | 84-99% | 95-102% | 88-103% | 
| Heparin plasma(n=5) | 86-95% | 91-106% | 87-100% | 95-101% |