| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.179 | 2.084 | 
| 250.00 | 1.612 | 1.517 | 
| 125.00 | 1.282 | 1.187 | 
| 62.50 | 0.933 | 0.838 | 
| 31.25 | 0.556 | 0.461 | 
| 15.63 | 0.297 | 0.202 | 
| 7.82 | 0.236 | 0.141 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-91% | 84% | 
| EDTA plasma(n=5) | 95-107% | 101% | 
| Heparin plasma(n=5) | 90-103% | 96% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-97% | 90-101% | 82-96% | 93-101% | 
| EDTA plasma(n=5) | 90-99% | 95-102% | 82-90% | 89-97% | 
| Heparin plasma(n=5) | 96-105% | 89-103% | 95-103% | 93-105% |