| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.201 | 2.114 |
| 1000.00 | 1.638 | 1.551 |
| 500.00 | 1.088 | 1.001 |
| 250.00 | 0.813 | 0.726 |
| 125.00 | 0.576 | 0.489 |
| 62.50 | 0.335 | 0.248 |
| 31.25 | 0.215 | 0.128 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-99% | 92% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-101% | 85-97% |
| EDTA plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |
| Heparin plasma(n=5) | 95-103% | 87-98% | 85-92% | 79-96% |