| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.163 | 2.054 | 
| 5.00 | 1.625 | 1.516 | 
| 2.50 | 1.053 | 0.944 | 
| 1.25 | 0.735 | 0.626 | 
| 0.63 | 0.532 | 0.423 | 
| 0.32 | 0.315 | 0.206 | 
| 0.16 | 0.184 | 0.075 | 
| 0.00 | 0.109 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-99% | 92% | 
| EDTA plasma(n=5) | 82-97% | 89% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-103% | 79-96% | 93-106% | 87-98% | 
| EDTA plasma(n=5) | 88-104% | 86-102% | 79-93% | 89-97% | 
| Heparin plasma(n=5) | 85-94% | 87-101% | 82-95% | 82-93% |