| 48 T | $370.00 |
| 96 T | $528.00 |
| 96 T*5 | $2244.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.164 | 2.052 |
| 500.00 | 1.625 | 1.513 |
| 250.00 | 1.135 | 1.023 |
| 125.00 | 0.886 | 0.774 |
| 62.50 | 0.543 | 0.431 |
| 31.25 | 0.388 | 0.276 |
| 15.63 | 0.234 | 0.122 |
| 0.00 | 0.112 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 82-96% | 87-103% | 95-102% |
| EDTA plasma(n=5) | 78-96% | 89-102% | 85-97% | 79-97% |
| Heparin plasma(n=5) | 92-104% | 87-96% | 89-101% | 82-90% |