| 64 T | $320.00 |
| 112 T | $458.00 |
| 112 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 80.00 | 2.678 | 2.589 |
| 40.00 | 1.648 | 1.559 |
| 20.00 | 1.012 | 0.923 |
| 10.00 | 0.534 | 0.445 |
| 5.00 | 0.302 | 0.213 |
| 2.50 | 0.191 | 0.102 |
| 1.25 | 0.133 | 0.044 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 89-96% | 93% |
| Heparin plasma(n=5) | 79-87% | 83% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-93% | 78-85% | 78-91% | 83-90% |
| EDTA plasma(n=5) | 92-103% | 85-94% | 87-95% | 87-95% |
| Heparin plasma(n=5) | 92-104% | 90-101% | 90-104% | 81-94% |