Products

Human HP(Haptoglobin) ELISA Kit
ELK1054
Size: Price:
48 T $265.00
96 T $376.00

OverviewPublications

Product name: Human HP(Haptoglobin) ELISA Kit
Reactivity: Human
Alternative Names: HP; Hp2-Alpha; Alpha-2-Macroglobulin; Zonulin; Hpt; Haptoglobin
Assay Type: Sandwich
Sensitivity: 2.87 ng/mL
Standard: 400 ng/mL
Detection Range: 6.25-400 ng/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Infection immunity;Hematology;
Uniprot ID: P00738
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human HP. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human HP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human HP, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human HP in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
400.00 2.119 1.967
200.00 1.896 1.744
100.00 1.649 1.497
50.00 1.005 0.853
25.00 0.634 0.482
12.50 0.428 0.276
6.25 0.238 0.086
0.00 0.152 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant HP and the recovery rates were calculated by comparing the measured value to the expected amount of HP in samples.
Matrix Recovery range Average
serum(n=5) 94-106% 100%
EDTA plasma(n=5) 80-97% 88%
Heparin plasma(n=5) 77-95% 86%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-102% 79-92% 87-103% 89-98%
EDTA plasma(n=5) 88-97% 91-98% 86-103% 79-91%
Heparin plasma(n=5) 85-104% 87-106% 83-92% 82-90%