| 64 T | $320.00 |
| 112 T | $458.00 |
| 112 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.511 | 2.436 |
| 5.00 | 1.829 | 1.754 |
| 2.50 | 1.130 | 1.055 |
| 1.25 | 0.575 | 0.500 |
| 0.63 | 0.261 | 0.186 |
| 0.32 | 0.189 | 0.114 |
| 0.16 | 0.141 | 0.066 |
| 0.00 | 0.075 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-92% | 89% |
| EDTA plasma(n=5) | 85-99% | 92% |
| Heparin plasma(n=5) | 85-96% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-98% | 95-106% | 90-102% | 77-83% |
| EDTA plasma(n=5) | 87-97% | 95-101% | 86-93% | 86-93% |
| Heparin plasma(n=5) | 92-102% | 92-106% | 85-96% | 82-90% |