Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.379 | 2.312 |
500.00 | 2.109 | 2.042 |
250.00 | 1.567 | 1.500 |
125.00 | 0.989 | 0.922 |
62.50 | 0.611 | 0.544 |
31.25 | 0.361 | 0.294 |
15.63 | 0.225 | 0.158 |
0.00 | 0.067 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 84-96% | 90% |
Heparin plasma(n=5) | 87-101% | 94% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 83-97% | 92-103% | 87-98% | 93-101% |
EDTA plasma(n=5) | 83-96% | 86-95% | 90-102% | 87-98% |
Heparin plasma(n=5) | 87-95% | 85-96% | 92-102% | 81-92% |