Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 1.941 | 1.838 |
1000.00 | 1.484 | 1.381 |
500.00 | 1.069 | 0.966 |
250.00 | 0.749 | 0.646 |
125.00 | 0.581 | 0.478 |
62.50 | 0.415 | 0.312 |
31.25 | 0.241 | 0.138 |
0.00 | 0.103 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 85-97% | 91% |
EDTA plasma(n=5) | 96-107% | 101% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-102% | 93-101% | 97-105% | 95-103% |
EDTA plasma(n=5) | 88-97% | 86-92% | 81-93% | 89-105% |
Heparin plasma(n=5) | 85-94% | 79-96% | 92-101% | 83-90% |