Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.027 | 1.941 |
500.00 | 1.700 | 1.614 |
250.00 | 1.243 | 1.157 |
125.00 | 0.828 | 0.742 |
62.50 | 0.508 | 0.422 |
31.25 | 0.340 | 0.254 |
15.63 | 0.174 | 0.088 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 81-97% | 89% |
EDTA plasma(n=5) | 85-99% | 92% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-102% | 82-96% | 87-98% | 78-92% |
EDTA plasma(n=5) | 97-105% | 86-98% | 84-93% | 89-102% |
Heparin plasma(n=5) | 85-94% | 87-105% | 92-101% | 79-90% |