Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.168 | 2.027 |
500.00 | 1.841 | 1.700 |
250.00 | 1.384 | 1.243 |
125.00 | 0.969 | 0.828 |
62.50 | 0.649 | 0.508 |
31.25 | 0.481 | 0.340 |
15.63 | 0.315 | 0.174 |
0.00 | 0.141 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-97% | 89% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 81-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-101% | 78-96% | 79-98% | 93-102% |
EDTA plasma(n=5) | 88-97% | 91-103% | 86-92% | 89-97% |
Heparin plasma(n=5) | 88-104% | 87-96% | 85-93% | 82-90% |