Product name: Mouse sCD40L(Soluble Cluster Of Differentiation 40 Ligand) ELISA Kit
Reactivity: Mouse
Alternative Names: CD154; TRAP; HIGM1; IGM; IMD3; TBAM; T-BAM; TNFSF5; Gp39; TNF Superfamily Member 5; Hyper-IgM Syndrome; TNF-Related Activation Protein; T-Cell B-Cell Activating Molecule
Assay Type: Sandwich
Sensitivity: 6.8 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates and other biological fluids
Assay Length: 3.5h
Research Area: Signal transduction;CD & Adhesion molecule;Apoptosis;Tumor immunity;Infection immunity;Immune molecule;
Uniprot ID: P27548
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse sCD40L. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse sCD40L. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse sCD40L, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse sCD40L in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
1000.00 2.035 1.939
500.00 1.576 1.480
250.00 1.184 1.088
125.00 0.831 0.735
62.50 0.527 0.431
31.25 0.336 0.240
15.63 0.152 0.056
0.00 0.096 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant sCD40L and the recovery rates were calculated by comparing the measured value to the expected amount of sCD40L in samples.
Matrix Recovery range Average
serum(n=5) 93-101% 97%
EDTA plasma(n=5) 84-97% 90%
Heparin plasma(n=5) 95-107% 101%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sCD40L and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-97% 87-96% 89-103% 82-95%
EDTA plasma(n=5) 81-97% 82-96% 95-102% 85-92%
Heparin plasma(n=5) 80-96% 93-105% 83-97% 95-104%