| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.075 | 1.941 | 
| 500.00 | 1.593 | 1.459 | 
| 250.00 | 1.185 | 1.051 | 
| 125.00 | 0.796 | 0.662 | 
| 62.50 | 0.538 | 0.404 | 
| 31.25 | 0.364 | 0.230 | 
| 15.63 | 0.256 | 0.122 | 
| 0.00 | 0.134 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-103% | 99% | 
| EDTA plasma(n=5) | 92-102% | 97% | 
| Heparin plasma(n=5) | 81-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 83-92% | 91-96% | 85-93% | 95-101% | 
| EDTA plasma(n=5) | 81-93% | 86-98% | 91-103% | 82-95% | 
| Heparin plasma(n=5) | 87-95% | 85-98% | 82-97% | 92-101% |