Overview

Product name: Human anti-HAV IgG(anti-Hepatitis A antibody IgG) ELISA Kit
Reactivity: Human
Alternative Names: Hepatitis A virus IgG
Assay Type: Sandwich
Sensitivity: 0.063 ng/mL
Standard: 10 ng/mL
Detection Range: 0.16-10 ng/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Infection immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human anti-HAV. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human anti-HAV. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human anti-HAV, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human anti-HAV in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
10.00 2.052 1.955
5.00 1.566 1.469
2.50 1.102 1.005
1.25 0.892 0.795
0.63 0.527 0.430
0.32 0.336 0.239
0.16 0.218 0.121
0.00 0.097 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant anti-HAV and the recovery rates were calculated by comparing the measured value to the expected amount of anti-HAV in samples.
Matrix Recovery range Average
serum(n=5) 83-96% 89%
EDTA plasma(n=5) 80-95% 87%
Heparin plasma(n=5) 90-105% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of anti-HAV and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-103% 95-103% 84-98% 93-102%
EDTA plasma(n=5) 87-98% 93-105% 89-97% 85-97%
Heparin plasma(n=5) 85-92% 82-96% 87-105% 83-92%