Products

Rat 25-OH-D(25 Hydroxy Vitamin D) ELISA Kit
ELK0843
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Rat 25-OH-D(25 Hydroxy Vitamin D) ELISA Kit
Reactivity: Rat
Alternative Names: 25 Hydroxy Vitamin D;
Assay Type: Competitive Inhibition
Sensitivity: 0.93 ng/mL
Standard: 200 ng/mL
Detection Range: 3.13-200 ng/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 2h
Research Area: Reproductive science;Genetic science;Nutrition metabolism;Bone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Rat 25-OH-D. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat 25-OH-D. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat 25-OH-D in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
200.00 0.179
100.00 0.357
50.00 0.596
25.00 0.812
12.50 1.197
6.25 1.583
3.13 1.789
0.00 2.211

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant 25-OH-D and the recovery rates were calculated by comparing the measured value to the expected amount of 25-OH-D in samples.
Matrix Recovery range Average
serum(n=5) 78-95% 86%
EDTA plasma(n=5) 82-95% 88%
Heparin plasma(n=5) 95-107% 101%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 25-OH-D and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 79-94% 87-96% 78-93% 82-90%
EDTA plasma(n=5) 95-102% 89-96% 87-98% 85-93%
Heparin plasma(n=5) 92-103% 93-105% 88-101% 97-106%