Size: | Price: |
---|---|
64 T | $320.00 |
112 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
100.00 | 2.195 | 2.124 |
50.00 | 1.602 | 1.531 |
25.00 | 1.216 | 1.145 |
12.50 | 0.741 | 0.670 |
6.25 | 0.541 | 0.470 |
3.13 | 0.315 | 0.244 |
1.57 | 0.232 | 0.161 |
0.00 | 0.071 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 85-97% | 91% |
EDTA plasma(n=5) | 78-92% | 85% |
Heparin plasma(n=5) | 92-105% | 98% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-102% | 87-96% | 78-92% | 85-97% |
EDTA plasma(n=5) | 98-105% | 86-95% | 94-103% | 95-102% |
Heparin plasma(n=5) | 95-103% | 88-96% | 81-95% | 91-102% |