| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 600.00 | 0.175 | |
| 300.00 | 0.389 | |
| 150.00 | 0.642 | |
| 75.00 | 0.856 | |
| 37.50 | 1.139 | |
| 18.75 | 1.596 | |
| 9.38 | 1.799 | |
| 0.00 | 2.132 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 88-102% | 95% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 93-101% | 85-98% | 90-101% |
| EDTA plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |
| Heparin plasma(n=5) | 95-103% | 79-96% | 93-106% | 87-98% |