48 T | $265.00 |
96 T | $376.00 |
Concentration (mIU/mL) | OD | Corrected OD |
---|---|---|
80.00 | 1.948 | 1.851 |
40.00 | 1.527 | 1.430 |
20.00 | 1.240 | 1.143 |
10.00 | 0.796 | 0.699 |
5.00 | 0.583 | 0.486 |
2.50 | 0.312 | 0.215 |
1.25 | 0.207 | 0.110 |
0.00 | 0.097 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 90-105% | 97% |
EDTA plasma(n=5) | 86-99% | 92% |
Heparin plasma(n=5) | 80-95% | 87% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |
EDTA plasma(n=5) | 81-94% | 93-107% | 93-102% | 91-105% |
Heparin plasma(n=5) | 92-101% | 82-98% | 89-97% | 85-93% |