Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
4000.00 | 2.266 | 2.170 |
2000.00 | 1.527 | 1.431 |
1000.00 | 1.251 | 1.155 |
500.00 | 0.865 | 0.769 |
250.00 | 0.453 | 0.357 |
125.00 | 0.312 | 0.216 |
62.50 | 0.206 | 0.110 |
0.00 | 0.096 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 90-105% | 97% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 83-95% | 89% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-97% | 87-96% | 82-95% | 92-105% |
EDTA plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |
Heparin plasma(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |