| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.266 | 2.183 |
| 2500.00 | 1.674 | 1.591 |
| 1250.00 | 1.243 | 1.160 |
| 625.00 | 0.845 | 0.762 |
| 312.50 | 0.585 | 0.502 |
| 156.25 | 0.293 | 0.210 |
| 78.13 | 0.179 | 0.096 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |
| EDTA plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |
| Heparin plasma(n=5) | 88-95% | 98-107% | 90-99% | 88-97% |
