Products

Mouse AMH(Anti-Mullerian Hormone) ELISA Kit
ELK0210
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Mouse AMH(Anti-Mullerian Hormone) ELISA Kit
Reactivity: Mouse
Alternative Names: MIF; MIH; MIS; Müllerian Inhibiting Factor; Müllerian Inhibiting Hormone; Müllerian Inhibiting Substance
Assay Type: Sandwich
Sensitivity: 0.95 ng/mL
Standard: 150 ng/mL
Detection Range: 2.35-150 ng/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Endocrinology;Reproductive science;Hormone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse AMH. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse AMH. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse AMH, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse AMH in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
150.00 2.218 2.129
75.00 1.525 1.436
37.50 1.183 1.094
18.75 0.857 0.768
9.38 0.597 0.508
4.69 0.361 0.272
2.35 0.179 0.090
0.00 0.089 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant AMH and the recovery rates were calculated by comparing the measured value to the expected amount of AMH in samples.
Matrix Recovery range Average
serum(n=5) 95-105% 100%
EDTA plasma(n=5) 93-101% 97%
Heparin plasma(n=5) 85-103% 94%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of AMH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-97% 96-105% 86-92% 89-97%
EDTA plasma(n=5) 88-101% 93-106% 85-95% 92-104%
Heparin plasma(n=5) 82-96% 95-102% 87-98% 81-97%