Products

Mouse IL18(Interleukin 18) ELISA Kit
ELK2269
Size: Price:
48 T $320.00
96 T $458.00

OverviewPublications

Product name: Mouse IL18(Interleukin 18) ELISA Kit
Reactivity: Mouse
Alternative Names: IL-18;IGIF; IL-1g; IL1F4; IL1-F4; Interferon-Gamma-Inducing Factor; Interleukin-1 Family Member 4; Iboctadekin; Interleukin-1 Gamma
Assay Type: Sandwich
Sensitivity: 37.1 pg/mL
Standard: 8000 pg/mL
Detection Range: 125-8000 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Infection immunity;
Uniprot ID: P70380
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse IL18. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse IL18. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse IL18, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse IL18 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
8000.00 1.941 1.853
4000.00 1.484 1.396
2000.00 1.036 0.948
1000.00 0.625 0.537
500.00 0.373 0.285
250.00 0.300 0.212
125.00 0.216 0.128
0.00 0.088 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant IL18 and the recovery rates were calculated by comparing the measured value to the expected amount of IL18 in samples.
Matrix Recovery range Average
serum(n=5) 91-99% 95%
EDTA plasma(n=5) 82-95% 88%
Heparin plasma(n=5) 83-98% 91%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL18 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-97% 90-98% 80-93% 92-105%
EDTA plasma(n=5) 87-98% 85-92% 79-96% 89-100%
Heparin plasma(n=5) 79-91% 89-103% 82-95% 86-99%